Reliability and Intersession Agreement of Microperimetric and Fixation Measurements Obtained with a New Microperimeter in Normal Eyes.

2015

Purpose: To evaluate the reliability and intersession agreement of measurements of retinal sensitivity as well as of the fixation pattern obtained in healthy eyes with a microperimeter integrating the mechanism of the scanning laser ophthalmoscope (SLO) with the static perimetry. Methods: This study included a sample of 44 healthy eyes of 44 subjects of a mean age of 27.0 ± 8.5 years. In all cases, microperimetric exams with the MAIA system (Centervue, Padova, Italy) were performed in three different sessions to evaluate the intersession repeatability. The consistency of measurements was analyzed by using the Friedman test and by analyzing the correlation between consecutive measurements. A…

MAIAAdultMalemedicine.medical_specialtyScanning laser ophthalmoscopeVisual AcuityFixation OcularSensitivity and SpecificityMicroperimetryRetinaMicroperimeterCellular and Molecular NeuroscienceYoung AdultOpticsOphthalmologymedicineHumansMathematicsÓpticabusiness.industryPreferred retinal locusReproducibility of ResultsMean ageRepeatabilityMiddle AgedFixationSensory SystemsHealthy VolunteersOphthalmologyFriedman testFixation (visual)Visual Field TestsFemaleVisual FieldsbusinessMicroperimetryRetinal locusCurrent eye research